에스지 하이테코
|
Performance Specifications Agilent 1260 Series Fluorescence Detector |
|
|
Detector type |
Multi-signal fluorescence detector with rapid on-line scanning capabilities and spectral data analysis |
|
Performance specifications |
Single wavelength operation: • RAMAN (H2O) > 500 (noise reference measured at signal) Ex=350 nm, Em=397 nm, dark value 450 nm, standard flow cell • RAMAN (H2O) > 3000 (noise reference measured at dark value) Ex=350 nm, Em=397 nm, dark value 450 nm, standard flow cell Dual wavelength operation: RAMAN (H2O) > 300 Ex 350 nm, Em 397 nm and Ex 350 nm, Em 450 nm, standard flow cell. |
|
Light source |
Xenon Flash Lamp, normal mode 20 W, economy mode 5 W, lifetime 4000 h |
|
Pulse frequency |
296 Hz for single signal mode 74 Hz for economy mode |
|
Excitation monochromator |
Range: settable 200 nm - 1200 nm and zero-order Bandwidth: 20 nm (fixed) Monochromator: concave holographic grating, F/1.6, blaze: 300 nm |
|
Drift |
< 0.9·10-3 AU/h at 254 nm |
|
Linear absorbance range |
> 2 AU (5 %) at 265 nm |
|
Wavelength accuracy |
± 1 nm |
|
Wavelength bunching |
1 – 400 nm |
|
Slit width |
1, 2, 4, 8, 16 nm |
|
Diode width |
< 1 nm |