Specifications Agilent 1200 Series Quaternary Pump
|
|
Flow precision
|
≤0.07 % RSD, or ≤0.02 min SD whatever is greater, based on retention time at constant room temperature
|
|
Flow accuracy
|
± 1 % or 10 µL/min whatever is greater
|
|
Range
|
Set points from 0.001 to 10.0 ml/min, in 0.001 ml/min increments
|
|
Operating range
|
0 - 40 MPa (0 - 400 bar, 0 - 5880 psi) up to 5 ml/min
0-20 MPa(0-200bar,0-2950psi) up to 10ml/min
|
|
Pulsation
|
<2 % amplitude (typically <1 %) at 1 mL/min isopropanol at all pressures >1 MPa (> 10 bar, >147 psi)
|
|
Compressibility compensation
|
User selectable, based on mobile phase compressibility
|
|
Delay volume
|
800 - 1100 µL depending on back pressure
|
|
Composition precision
|
<0.20 % SD, at 0.2 and 1 mL/min
|
|
Recommended pH range
|
1.0 - 12.5
Solvent swith pH < 2.3 should not contain acid swhich attack steel
|